Many people pay attention to the maintenance of oxygen generator and teach you some tips to increase the service life of your oxygen generator by at least one to two years
1、 Before starting the machine, read the manual carefully to make sure that the filter cotton or filter element has been installed on the machine, and the machine can be used when it is not installed.
2、 The oxygen generator should be placed stably, otherwise it will increase the noise during the operation of the oxygen generator. In the process of transportation and storage, it is strictly prohibited to put it horizontally, upside down, wet or direct sunlight! At present, according to its manufacturing process, the general molecular sieve oxygen generator is not allowed to be inverted. Because there is a compressor in the oxygen machine, the compressor base is fixed by spring. Once inverted, the compressor will be adversely affected.
3、 Household oxygen inhalers should be protected from moisture. The oxygen machine should be placed in a dry and ventilated place.
4、 Household molecular sieve oxygen generator should be cleaned and replaced regularly. Air filter cotton: clean every week, and replace new filter element if necessary. It is different according to different environment. It is about 1000 hours in dusty, smoky and seriously polluted environment, and 1500 hours in normal condition.
5、 Disinfect and clean the oxygen tube, soak it in medical alcohol for 30 minutes, then rinse it with clean water and dry it.
6、 The water level in the humidification bottle should not be too high (the water level should be half of the bottle body), otherwise the water in the bottle will easily overflow or enter the oxygen absorption pipe. The humidification cup (bottle) of household oxygen generator is cleaned with distilled water, pure water or cold boiled water every day. After washing with soapy water, rinse with 1:10 vinegar solution, and then rinse thoroughly with hot water.
7、 The air inlet of the oxygen machine should not have any obstructions. To avoid blocking the entry of air.
8、 Do not adjust the flow meter of the oxygen generator at the zero position, so that the oxygen will not come out.
9、 When using the atomizer of oxygen generator for atomization, pay attention not to fill up the medicine in the atomizer cup, subject to placing 1 / 3.
10、 When the oxygen generator is not used for a long time, please cut off the power supply, pour out the water in the humidification bottle, wipe the surface of the oxygen generator, cover it with plastic cover, and store it in a dry place without sunlight. Dehumidification for a period of time.
11、 Oxygen concentration test should be carried out regularly.